McLane Research Laboratories, Inc.
About McLane Research LabsMcLane Research Labs ProductsMcLane Research Labs ServicesContact McLane Research LabsMcLane Research Labs Announcements
Matching Flow Rate to Filter Type

Sample integrity and pumping accuracy depend on properly matching flow rate to filter type. Based on the laboratory and field experience gained during our WTS-LV research and development effort, McLane recommends the pairings shown in the tables below.

If your research requires the use of a particular filter type, the information in the tables will act as a preliminary guide in selecting an appropriately sized LV pump. Our engineers are available to assist you with additional questions.

This compatibility table is also available as a PDF.

Matching Flow Rate to Filter Type - MRL Recommended Pairings
Pump
Designation
Flow Rate
Range [L/min]
Filter
Type
LV04 1 - 4membrane filters, Nucleopore
LV08 5 - 10glass fiber filters, GFF, QMA
LV2010 - 20mesh filters, Nytex
LV3015 - 30mesh filters, Nytex
LV5025 - 50mesh filters, Nytex

Filter Characteristics and Flow Rate Limits
Filter Type Pore Size
[microns]
Maximum
Flow Rate
[ml/min]
Polycarbonate and polyester 0.2 3,000
membrane filters 0.4 5,000
e.g., Nucleopore 0.8 8,000
  1.010,000
Glass fiber filters 0.8 8,000
e.g., GF/F, QMA 1.710,000
  5.020,000
Mesh filters, e.g., Nytex60.050,000

Note: These recommendations and limits are for filter diameter of 142 mm, the standard size for WTS-LV pumps. Larger pore sizes than those shown will permit higher pumping rates.

Home | About Us | Products | Services | News | Support | Contact/Reply
McLane Research Laboratories, Inc Falmouth Technology Park
121 Bernard E. Saint Jean Drive East Falmouth, MA 02536
Tel: 508.495.4000   Fax: 508.495.3333